Why more measurements can still be wrong
Does repeating a measurement remove every kind of error?
Entering midway? Revisit Does the treatment work, or did the groups differ? for the preceding worked example, or start here and use the explanation below.
Start with a concrete question
Original synthetic instrument: measure a known 10 cm reference with a fixed offset and a repeating disturbance pattern.
Try the initial inputs, then write a prediction before changing one setting. All calculations run locally on your CPU.
What do you expect?
Predict the raw mean after four readings with a +2 cm offset. Will twelve readings fix it?
Change one thing. Inspect what happens.
- Compare 1, 2, 4 and 12 readings while keeping the offset fixed.
- Set disturbance to zero. Explain why perfectly repeatable readings can be inaccurate.
- Use the reference calibration and compare raw and corrected means.
- Propose a second reference length to check whether one offset correction is sufficient for a different instrument.
Your settingsMechanism
Calculation belowEvidence
Compare outcomes
Compare the current results
Every bar starts at zero and uses the same numeric scale in this chart. The upper bound automatically adjusts to the largest current value (at least 1), so compare the numbers when changing inputs. Units and quantities come from each label; different kinds of quantity are not interchangeable. Displayed values are rounded to at most six significant digits.Follow the calculation
- Reading 1: 10 + bias 2 + disturbance -2.00 = 10.00 cm.
- Reading 2: 10 + bias 2 + disturbance 2.00 = 14.00 cm.
- Reading 3: 10 + bias 2 + disturbance -1.00 = 11.00 cm.
- Reading 4: 10 + bias 2 + disturbance 1.00 = 13.00 cm.
- Mean = sum 48.00 / 4 = 12.00 cm.
- Reference calibration removes known bias: 12.00 - (2) = 10.00 cm.
Repeating a biased instrument does not remove its offset. This fixed disturbance pattern is a controlled example, not a random sample or confidence interval.
Captures keep your inputs, prediction and observation. Results are recalculated from those inputs. Resetting inputs keeps your captures and challenge history.
Explain the mechanism
What the experiment explains
Original synthetic instrument: measure a known 10 cm reference with a fixed offset and a repeating disturbance pattern.
Precision describes repeatability; accuracy describes closeness to the reference. An instrument can be precise and consistently wrong.
The default four readings are 10, 14, 11, 13 cm. Their mean is 48 / 4 = 12 cm. The disturbances cancel but the +2 cm offset remains.
Subtracting a known offset gives 10 cm. Real calibration estimates an offset with uncertainty; it can also reveal scale errors that this model does not include.
Read the relationship
Reading = reference + bias + disturbance. Mean = sum of readings / reading count. Corrected mean = mean − known bias.
Use the trace as evidence
The trace recomputes from your current controls. Change one control at a time, record what changed, and compare the result with your prediction. The bars use the same quantities as the numerical trace; a larger bar is not automatically a better outcome.
Boundaries of the model
- The disturbance pattern −1, +1, −0.5, +0.5 repeats; it is not independent random noise.
- The reference and offset are known exactly in this teaching model; no confidence interval is implied.
Continue with Does the treatment work, or did the groups differ? to investigate a different mechanism.
Try a fresh case
A new 20 cm reference gives readings 22, 24, 23, 23 cm. A calibration establishes a +3 cm offset. What is the corrected mean in cm?
This is a public practice check. Retries support learning; matching a number does not certify independent proficiency.
Take the investigation further
Choose any extension. These are project prompts for you to carry out; they are not automatically assessed.
School investigation
Compare repeated ruler measurements and describe resolution separately from accuracy.
College investigation
Add random noise and estimate uncertainty without treating repeated bias as random error.
Workplace investigation
Write an instrument acceptance check with reference standards, tolerance, and recalibration triggers.
Keep the evidence, choose your next step
Completion records your own decision that you have worked through this investigation. Practice checks and independent project evidence are separate.